• Resuscitation · Oct 2005

    Comparative Study

    A comparison of three neonatal resuscitation devices.

    • Stacie Bennett, Neil N Finer, Wade Rich, and Yvonne Vaucher.
    • Department of Pediatrics, Division of Neonatology, UCSD Medical Center, 401 Dickenson, MPF 1-140, San Diego, CA 92103-8774, USA. scbennett@ucsd.edu
    • Resuscitation. 2005 Oct 1;67(1):113-8.

    BackgroundVentilation during neonatal resuscitation involves the use of self-inflating bags, flow-inflating bags, and T-piece resuscitators. The ability of operators to deliver desired peak inspiratory pressures (PIP), positive end expiratory pressures (PEEP), prolonged inflations and the length of time to transition between different pressures has not been compared for all three of these devices.ObjectiveTo compare the ability of neonatal resuscitation personnel to deliver predetermined ventilation interventions using these devices in advance of a clinical trial of neonatal resuscitation.Design/MethodsWe studied 31 operators (neomatologists, neonatal respiratory therapists, neonatal fellows, a pediatrician, pediatric residents, neonatal nurse practitioners, and neonatal nurses) using a T-piece resuscitator (Neopuff), Fisher and Paykel Healthcare, Auckland, New Zealand), a self-inflating bag (Baby Blue II, Vital Signs, Totowa, NJ), and a flow-inflating bag (Model E191 Anesthesia Associates, San Marcos, CA). The self-inflating bag was tested with and without the manufacturer's PEEP valve. Using a continuous pressure recording system and a neonatal manikin, we evaluated the ability to deliver a consistent PIP of 20 or 40 cmH2O and a PEEP of 5 cmH2O during 30 s of ventilation, the ability to maintain a 5 s inflation at a PIP of 20 cmH2O and the time to transition from a PIP of 20 to 40 cmH2O. Each device was evaluated with and without a qualitative CO2 detector (Pedicap) Nellcor Pleasanton, CA).ResultsThe T-piece resuscitator delivered the desired PIP more precisely and consistently compared with the self-inflating bag at a target of 20 cmH2O (maximum PIP 20.7 cmH2O, S.D.=0.8 versus 24.7 cmH2O, S.D.=2.8; p<0.001). At a target of 40 cmH2O, the maximum pressure delivered with the T-piece resuscitator was significantly less than both the flow-inflating bag and the self-inflating bag (39.7 cmH2O, S.D.=2.1 versus 44 cmH2O, S.D.=3.3 versus 45.3 cmH2O, S.D.=4.7; p<0.001). It took significantly longer to increase the PIP from 20 to 40 cmH2O using the T-piece resuscitator compared to the self-inflating bag or the flow-inflating bag (5.7 s versus 2.2 s versus 1.8 s; p<0.001), and three operators could not make the transition in the allotted 15 s time limit. During the 5 s prolonged inflation, the T-piece resuscitator and the flow-inflating bag maintained a pressure greater than 18 cmH2O for a longer time than the self-inflating bag (4 s versus 3.7 s versus 2.2 s; p<0.001). The self-inflating bag with the PEEP valve in place provided significantly less PEEP than both the T-piece resuscitator and the flow-inflating bag (3.6 cmH2O versus 4.4 cmH2O versus 4.4 cmH2O; p<0.005). The Pedicap did not significantly affect any of the observed results, and there were no consistent operator differences between different disciplines or years of experience.ConclusionsThe T-piece resuscitator delivered the desired pressures more accurately, but required greater time to increase the PIP from 20 to 40 cmH2O. It was difficult to maintain a prolonged inflation and deliver the desired PEEP with the self-inflating bag even with the PEEP valve in place. There is a need for improvement in the design and function of current manual resuscitation devices and for prospective trials to evaluate the optimal method of bag and mask ventilation during resuscitation of the newborn infant.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…