• Critical care medicine · May 1995

    Comparative Study

    Continuous versus bolus thermodilution cardiac output measurements--a comparative study.

    • T Mihaljevic, L K von Segesser, M Tönz, B Leskosek, B Seifert, R Jenni, and M Turina.
    • Department of Internal Medicine, University Hospital Zurich, Switzerland.
    • Crit. Care Med. 1995 May 1;23(5):944-9.

    ObjectiveTo compare the methods for continuous and bolus thermodilution cardiac output measurements.DesignIn vivo and in vitro experimental studies.SettingSurgical research division in a university hospital.SubjectsEight calves and flow bench model.InterventionsData were collected in vivo from eight calves instrumented with pulmonary artery catheters, which allowed both continuous and bolus thermodilution measurements. The pulmonary artery catheter was placed through the external jugular vein. All in vitro measurements were performed using a flow bench model.Measurements And Main ResultsA total of 232 bolus and continuous thermodilution measurements were analysed in vivo to determine the degree of agreement between the two methods. The absolute measurement bias was 0.14 L/min with 95% confidence limits ranging from -0.83 to 1.15 L/min. In vitro analysis of 576 measurements at six different temperature points (range 31 degrees to 41 degrees C), using clinically relevant flows (2 to 9 L/min), showed overestimation of flow values using continuous and bolus thermodilution methods. However, the continuous method showed better accuracy by a lower degree of overestimation. Systematic error was 9.7 +/- 8.4 (SD) % for continuous and 11.1 +/- 6.3% for the bolus method (p < .001). This effect was especially evident at lower flow rates. The influence of various temperatures on the accuracy and reproducibility of both methods of measurement was statistically significant but not clinically relevant. The infusion of lactated Ringer's lactate solution (infusion rates 100 to 1000 mL/hr) affects both methods at a low flow rate of 2 L/min, without causing a significant effect on continuous measurement at a higher flow rate (4 L/min). Shunting of 50% of circulating volume to the distal part of the thermal filament of the pulmonary catheter impaired the accuracy of continuous measurement without affecting results from bolus measurements (systematic error -26.8 +/- 8.2% for continuous and -5.2 +/- 4.1% for bolus thermodilution).ConclusionsContinuous thermodilution cardiac output measurement provided higher accuracy and greater resistance to thermal noise than standard bolus measurements. The correct placement of the catheter is essential for precise measurements.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…