Case Reports
[Anesthetic machine leakage from vaporizer by external force derived from keyboard of electronic medical records].
- Hiromi Ikegami, Ryokichi Goto, Syotarou Sakamoto, and Hanako Kohama.
- Department of Anesthesia, Yoka Hospital, Yabu 667-8555.
- Masui. 2012 Nov 1;61(11):1288-90.
AbstractWe experienced the leakage from the vaporizer of the anesthetic machine despite the normalities on performing the initial leak test. The vaporizer of the anesthetic machine was compressed by computer keyboard of EMR which caused a leak from vaporizer. After computer keyboard and the vaporizer were set at normal position, the leak stopped.