• Masui · Nov 2012

    Case Reports

    [Anesthetic machine leakage from vaporizer by external force derived from keyboard of electronic medical records].

    • Hiromi Ikegami, Ryokichi Goto, Syotarou Sakamoto, and Hanako Kohama.
    • Department of Anesthesia, Yoka Hospital, Yabu 667-8555.
    • Masui. 2012 Nov 1;61(11):1288-90.

    AbstractWe experienced the leakage from the vaporizer of the anesthetic machine despite the normalities on performing the initial leak test. The vaporizer of the anesthetic machine was compressed by computer keyboard of EMR which caused a leak from vaporizer. After computer keyboard and the vaporizer were set at normal position, the leak stopped.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…