• Crit Care · Jan 2012

    Comparative Study

    Comparison of accuracy of optic nerve ultrasound for the detection of intracranial hypertension in the setting of acutely fluctuating vs stable intracranial pressure: post-hoc analysis of data from a prospective, blinded single center study.

    • Venkatakrishna Rajajee, Jeffrey James Fletcher, Lauryn Renee Rochlen, and Teresa Lee Jacobs.
    • Crit Care. 2012 Jan 1;16(3):R79.

    IntroductionOptic nerve sheath diameter (ONSD) measurement with bedside ultrasound has been shown in many studies to accurately detect high intracranial pressure (ICP). The accuracy of point-in-time ONSD measurement in the presence of ongoing fluctuation of ICP between high and normal is not known. Recent laboratory investigation suggests that reversal of optic nerve sheath distension may be impaired following bouts of intracranial hypertension. Our objective was to compare the accuracy of ONSD measurement in the setting of fluctuating versus stable ICP.MethodsThis was a retrospective analysis of data from prospective study comparing ONSD to invasive ICP. Patients with invasive ICP monitors in the ICU underwent ONSD measurement with simultaneous blinded recording of ICP from the invasive monitor. Three measurements were made in each eye. Significant acute ICP fluctuation (SAIF) was defined in two different ways; as the presence of ICP both above and below 20 mmHg within a cluster of six measurements (Definition 1) and as a magnitude of fluctuation >10 mmHg within the cluster (Definition 2). The accuracy of point-in-time ONSD measurements for the detection of concurrent ICP >20 mmHg within clusters fulfilling a specific definition of SAIF was compared to the accuracy of ONSD measurements within clusters not fulfilling the particular definition by comparison of independent receiver operating characteristic (ROC) curves.ResultsA total of 613 concurrent ONSD-ICP measurements in 109 clusters were made in 73 patients. Twenty-three (21%) clusters fulfilled SAIF Definition 1 and 17 (16%) SAIF Definition 2. For Definition 1, the difference in the area under the curve (AUC) of ROC curves for groups with and without fluctuation was 0.10 (P = 0.0001). There was a fall in the specificity from 98% (95% CI 96 to 99%) to 74% (63 to 83%) and in the positive predictive value from 89% (80 to 95%) to 76% (66 to 84%) with fluctuation. For Definition 2, also, there was a significant difference between the AUC of ROC curves of groups with fluctuation-magnitude >10 mmHg and those with fluctuation-magnitude 5 to 10 mmHg (0.06, P = 0.04) as well as <5 mmHg (0.07, P = 0.01).ConclusionsSpecificity and PPV of ONSD for ICP >20 mmHg are substantially decreased in patients demonstrating acute fluctuation of ICP between high and normal. This may be because of delayed reversal of nerve sheath distension.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.