• EuroIntervention · Dec 2014

    Observational Study

    Emergent cardiac surgery during transcatheter aortic valve implantation (TAVI): insights from the Edwards SAPIEN Aortic Bioprosthesis European Outcome (SOURCE) registry.

    • Holger Eggebrecht, Rajendra H Mehta, Philipp Kahlert, Gerhard Schymik, Thierry Lefèvre, Rüdiger Lange, Carlos Macaya, Lazar Mandinov, Olaf Wendler, and Martyn Thomas.
    • Cardioangiological Center Bethanien (CCB), Frankfurt, Germany.
    • EuroIntervention. 2014 Dec 1;10(8):975-81.

    AimsTranscatheter aortic valve implantation (TAVI) carries the risk of intraprocedural complications that may ultimately require emergent cardiac surgery (ECS). However, few data exist on the incidence, reasons and outcomes of patients needing ECS during TAVI. We analysed data from 2,307 TAVI patients, prospectively enrolled in the multicentre Edwards SAPIEN Aortic Bioprosthesis European Outcome (SOURCE) registry.Methods And ResultsTwenty-seven (1.2%) of 2,307 patients required ECS. The rates of ECS were similar for patients undergoing transapical TAVI compared with transfemoral TAVI (1.1% vs. 1.2%). The leading causes for ECS were embolisation/migration of the TAVI valve prosthesis (9/27, 33%) and procedure-related aortic injury (n=7, 26%). Thirty-day mortality of ECS was high (51.9%) and showed cause-specific differences, with 100% mortality in patients with aortic rupture or cardiac tamponade, 0% death in those with acute aortic regurgitation and intermediate risk of death or intermediate mortality in those with aortic injury or valve embolisation/migration.ConclusionsRates of ECS during TAVI were low (1.2%). Although ECS was performed without time delay, emergent surgery was associated with a 30-day mortality of 52%. Complications with dramatic acute consequences (annular rupture, aortic injury) had higher mortality than those with less acute deterioration (aortic regurgitation). Prevention of complications requiring ECS during TAVI appears to be of critical importance, focusing on less traumatic, more flexible delivery catheter systems and retrievable valves to reduce the risk of aortic injury and valve embolisation, the two most common causes of ECS.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.