• Crit Care · Jan 2012

    Effects of respiratory mechanics on the capnogram phases: importance of dynamic compliance of the respiratory system.

    • Barna Babik, Zsófia Csorba, Dorottya Czövek, Patrick N Mayr, Gábor Bogáts, and Ferenc Peták.
    • Crit Care. 2012 Jan 1;16(5):R177.

    IntroductionThe slope of phase III of the capnogram (SIII) relates to progressive emptying of the alveoli, a ventilation/perfusion mismatch, and ventilation inhomogeneity. S(III) depends not only on the airway geometry, but also on the dynamic respiratory compliance (Crs); this latter effect has not been evaluated. Accordingly, we established the value of SIII for monitoring airway resistance during mechanical ventilation.MethodsSidestream capnography was performed during mechanical ventilation in patients undergoing elective cardiac surgery (n = 144). The airway resistance (Raw), total respiratory resistance and Crs displayed by the ventilator, the partial pressure of arterial oxygen (PaO2) and S(III) were measured in time domain (S(T-III)) and in a smaller cohort (n = 68) by volumetry (S(V-III)) with and without normalization to the average CO2 phase III concentration. Measurements were performed at positive end-expiratory pressure (PEEP) levels of 3, 6 and 9 cmH2O in patients with healthy lungs (Group HL), and in patients with respiratory symptoms involving low (Group LC), medium (Group MC) or high Crs (Group HC).ResultsS(T-III) and S(V-III) exhibited similar PEEP dependencies and distribution between the protocol groups formed on the basis of Crs. A wide interindividual scatter was observed in the overall Raw-S(T-III) relationship, which was primarily affected by Crs. Decreases in Raw with increasing PEEP were reflected in sharp falls in S(III) in Group HC, and in moderate decreases in S(III) in Group MC, whereas S(T-III) was insensitive to changes in airway caliber in Groups LC and HL.ConclusionsSIII assessed in the time domain and by volumetry provide meaningful information about alterations in airway caliber, but only within an individual patient. Although S(T-III) may be of value for bedside monitoring of the airway properties, its sensitivity depends on Crs. Thus, assessment of the capnogram shape should always be coupled with Crs when the airway resistance or oxygenation are evaluated.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…