J. Cardiothorac. Vasc. Anesth. · Apr 2000
Clinical TrialIs the placement of a pulmonary artery catheter still justified solely for the measurement of cardiac output?
- S G Sakka, K Reinhart, K Wegscheider, and A Meier-Hellmann.
- Department of Anesthesiology and Intensive Care Medicine, Friedrich-Schiller-University of Jena, Germany.
- J. Cardiothorac. Vasc. Anesth. 2000 Apr 1;14(2):119-24.
ObjectiveThe authors compared four clinical techniques of measuring cardiac output (CO) in critically ill patients: pulmonary artery thermodilution (CO[PA]), transpulmonary aortic thermodilution (CO[AORTA]), Fick principle-derived (CO[FICK]), and continuous pulmonary artery (CCO) measurements.DesignProspective clinical study.SettingSurgical intensive care unit of a university hospital.ParticipantsTwelve adult patients suffering from sepsis or septic shock.InterventionsAll patients were deeply sedated and mechanically ventilated in a pressure-controlled mode. Each patient received a 7.5F five-lumen pulmonary artery catheter for the continuous measurement of cardiac output and a 4F aortic catheter with an integrated thermistor. The thermistors of the two different catheters were connected to one computer system (COLD-Z021, Pulsion Medical Systems, Munich, Germany). Whole-body oxygen consumption was measured by indirect calorimetry using a metabolic cart (Deltatrac, Datex-Engstroem, Helsinki, Finland) over a 5-minute period, at the end of which arterial and mixed venous blood gases were taken and measured by co-oximetry. During each measuring period, three bolus CO measurements were performed. A total number of 51 CO measurements was analyzed.ResultsLinear regression analysis revealed the highest correlation between CO(AORTA) and CO(PA) (r = 0.98), whereas agreement between these two techniques and CCO was lower (r = 0.92 and r = 0.93). All three techniques correlated comparably with CO(FICK) (r = 0.85, r = 0.83, and r = 0.83).ConclusionThe correlations among the four CO techniques were high and similar, with CO(PA) and CO(AORTA) techniques showing the highest agreement. Because CO with similar accuracy can be obtained from transpulmonary aortic thermodilution in a less-invasive manner, it appears that the placement of a pulmonary artery catheter solely for the measurement of CO is no longer justified, unless continuous CO measurements are needed.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.