    Use of pulse oximetry to monitor venous saturation during extracorporeal life support.

    • R M Odell, R M Doran, R C Erickson, T C Zeilinger, and M G Coyle.
    • Department of Neonatology, Huntington Memorial Hospital, Pasadena, CA 91109-7013.
    • Crit. Care Med. 1994 Apr 1;22(4):626-32.

    ObjectiveTo assess the ability of two different pulse oximeters to display continuous venous oxygen saturation through an extracorporeal bypass circuit with a degree of accuracy comparable to direct in-line oximetry.DesignProspective, comparison study of pulse oximeters (test oximeter 1 or test oximeter 2) and an in-line oximeter (test oximeter 3).SettingA tertiary care neonatal intensive care unit.PatientsSixty-five consecutive neonates with severe cardiorespiratory failure undergoing extracorporeal life support.InterventionsThe accuracy of the oximeters was determined by simultaneously comparing the saturation displayed by the pulse oximeters (test oximeters 1 and 2) and/or the in-line oximeter (test oximeter 3) with the measured fractional venous oxygen saturation obtained at regular intervals from the extracorporeal circuit.Main Outcome MeasuresVenous oxygen saturation was the criterion standard used to determine accuracy. Bias was defined as the mean difference between observed pulse oximeter or in-line oximeter values and the measured venous oxygen saturation. Mean biases were calculated for venous oxygen saturation measurements between 55% and 99% at intervals of 10%. Precision (the standard deviation of the bias) was calculated for low (55% to 75%), medium (76% to 81%), and high (82% to 99%) venous oxygen saturation values. A total of 983 venous oxygen saturation measurements were made and compared with simultaneous oximeter readings from test oximeter 1 (n = 600), test oximeter 2 (n = 478), and test oximeter 3 (n = 587).ResultsTest oximeter 1 was the most precise instrument at each level of venous oxygen saturation (SD, 4.0 to 4.8). Test oximeter 3 demonstrated the most consistent mean bias (range, 8), but was the most inaccurate oximeter across all levels of venous oxygen saturation.ConclusionsIn addition to its known clinical usefulness, pulse oximetry may serve as an adequate substitute for in-line oximetry during extracorporeal life support.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.