-
Anesthesia and analgesia · Nov 2005
Comparative StudyAn in vitro comparison of the electrical conducting properties of multiport versus single-port epidural catheters for the epidural stimulation test.
- Ban C H Tsui and Corey K C Sze.
- Department of Anesthesiology and Pain Medicine, 8-120 Clinical Sciences Building, University of Alberta, Edmonton, Alberta, Canada T6G 2G3. btsui@ualberta.ca
- Anesth. Analg. 2005 Nov 1;101(5):1528-30.
AbstractEffective conduction of electricity through a catheter is essential for the success of the epidural stimulation test. In this in vitro study we examined the electrical conductivity of single and multiport epidural catheters (with and without embedded metal elements) after being primed with normal saline. Seven different types of 19-gauge catheters (n = 5), either single-port or multiport catheters, with or without embedded metal elements, were studied. The proximal end of each epidural catheter was connected to the cathode of a nerve stimulator via an electrode adapter. The catheter, primed with normal saline, was placed at the bottom of a syringe filled with 5 different volumes of saline (1, 2, 3, 4 and 5 mL) and attached to an electrode adapter. The voltage of the peripheral nerve stimulator was measured using an oscilloscope. The electrical resistance between the proximal and distal end of the catheter was calculated using Ohm's Law. In catheters without metal elements the electrical resistances were too high to be measured. In catheters that had metal elements, the mean electrical resistances of the same catheter design (single-port or multiport) were similar. However, the electrical resistances of the multiport metal reinforced epidural catheters were significantly lower (P < 0.05) than the single-port metal coil reinforced epidural catheters. The volume of saline in the syringe had no impact on the measured electrical resistances. This study suggests that multiport metal reinforced epidural catheters have low electrical resistances and, thus, are a reasonable alternative to single-port catheters for transmitting sufficient current for performing the epidural stimulation test. On the other hand, epidural catheters without metal elements (single-port or multiport) are not suitable for performing the stimulation test.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.