    Noninvasive continuous blood pressure measurement during anaesthesia: a clinical evaluation of a method commonly used in measuring devices.

    • J R De Jong, H H Ros, and J J De Lange.
    • Free University Hospital, Department of Anaesthesiology, Amsterdam, The Netherlands.
    • Int J Clin Monit Comput. 1995 Feb 1;12(1):1-10.

    AbstractThe objective of the study was assess the utility during anaesthesia of noninvasive continuous blood pressure measurement techniques which use intermittent oscillometric blood pressure measurement for their calibration. The assessment was performed by comparing noninvasive blood pressure with intra-arterial blood pressure. The noninvasive blood pressure measurement device used for evaluation was the NCAT N-500 which uses tonometry for its continuous measurements. Fifteen patients were studied. In 10 patients the intra-arterial blood pressure curve (IBP) was recorded from the radial artery (radial artery group), and in 5 patients it was recorded from the brachial artery (brachial artery group). In all patients the oscillometrically calibrated tonometric blood pressure (OTBP) was recorded from the other arm. To discriminate between calibration dependent measurement error and tonometric measurement error, the OTBP signal was recalibrated against the IBP signal to get the intra-arterial calibrated tonometric pressure curve (ITBP). OTBP-IBP reflected the overall measurement error, ITBP-IBP the error of the tonometric measurement, and OTBP-ITBP the calibration dependent measurement error. According to criteria formulated in the discussion the accuracy and agreement of the ITBP-IBP measurements were clinical acceptable. Accuracy and agreement of OTBP-IBP and of OTBP-ITBP were not clinical acceptable. Correlation of dynamic behavior was lower for OTBP than for ITBP. A significant effect of site difference between calibration measurements and continuous measurements was not found. It is concluded that the approach of continuous noninvasive blood pressure measurement based on the combination of two different measurement methods, in which the continuous method is calibrated by the oscillometric method, lead to clinical unacceptable accuracy and agreement in the patient group studied.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.