Anesthesia and analgesia · Oct 2009
Impact of multiaccess infusion devices on in vitro drug delivery during multi-infusion therapy.
- Bertrand Décaudin, Sophie Dewulf, Damien Lannoy, Nicolas Simon, Alexandre Secq, Christine Barthélémy, Bertrand Debaene, and Pascal Odou.
- Department of Biopharmacy, Galenic and Hospital Pharmacy EA 4034-IFR114, Lille 2 University, Lille, France.
- Anesth. Analg. 2009 Oct 1;109(4):1147-55.
BackgroundMultiaccess infusion sets allow multiple simultaneous infusions but may induce interference in drug delivery resulting from large variations in the delivery rate of potent drugs. In this study, we sought to understand the influence of multiaccess infusion device properties (dead space volume and antireflux valve [ARV]) on drug delivery during multi-infusion therapy.MethodsInfusion sets differing in length, dead space volume, and presence of an ARV were assessed. Three drugs were infused simultaneously through different access points, and their concentrations were obtained using UV spectrophotometric analysis of the effluent. Different infusion configurations were compared by assessing (1) the amount of drug delivered to the patient per unit of time, (2) the mean amount of drug delivered to the patient per unit of time during the steady-state infusion (mass flow rate plateau), and (3) flow change efficiency calculated from the ratio of the area under the experimental instant mass flow rate curve to the area corresponding to theoretical instant mass flow rate curve.ResultsInfusion sets with lower dead space volumes offered significantly higher flow change efficiency (53.0% +/- 15.4% with a dead space volume equal to 0.046 mL 5 min after the start of infusion) than infusion sets with higher dead space volume (5.6% +/- 8.2% with a dead space volume equal to 6.16 mL), whatever the flow rate changes. Even in case of large dead space volumes, the presence of an ARV significantly increased the mass flow rate plateau (from 92.4% to 99.3% of the theoretical plateau without and with the presence of an ARV, respectively).ConclusionsMulti-infusion therapy induces perturbation in drug delivery. These perturbations (lag time, backflow, and bolus) could be reduced by using infusion sets including very low dead space volume and an ARV.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.