J Clin Monit Comput · Feb 2009
Specificity improvement for network distributed physiologic alarms based on a simple deterministic reactive intelligent agent in the critical care environment.
- James M Blum, Grant H Kruger, Kathryn L Sanders, Jorge Gutierrez, and Andrew L Rosenberg.
- Department of Anesthesiology and Critical Care, The University of Michigan Health Systems, 4172 Cardiovascular Center/SPC 5861, 1500 East Medical Center Drive, Ann Arbor, MI 48109-5861, USA. jmblum@umich.edu
- J Clin Monit Comput. 2009 Feb 1;23(1):21-30.
AbstractAutomated physiologic alarms are available in most commercial physiologic monitors. However, due to the variability of data coming from the physiologic sensors describing the state of patients, false positive alarms frequently occur. Each alarm requires review and documentation, which consumes clinicians' time, may reduce patient safety through 'alert fatigue' and makes automated physician paging infeasible. To address these issues a computerized architecture based on simple reactive intelligent agent technology has been developed and implemented in a live critical care unit to facilitate the investigation of deterministic algorithms for the improvement of the sensitivity and specificity of physiologic alarms. The initial proposed algorithm uses a combination of median filters and production rules to make decisions about what alarms to generate. The alarms are used to classify the state of patients and alerts can be easily viewed and distributed using standard network, SQL database and Internet technologies. To evaluate the proposed algorithm, a 28 day study was conducted in the University of Michigan Medical Center's 14 bed Cardiothoracic Intensive Care Unit. Alarms generated by patient monitors, the intelligent agent and alerts documented on patient flow sheets were compared. Significant improvements in the specificity of the physiologic alarms based on systolic and mean blood pressure was found on average to be 99% and 88% respectively. Even through significant improvements were noted based on this algorithm much work still needs to be done to ensure the sensitivity of alarms and methods to handle spurious sensor data due to patient or sensor movement and other influences.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.