• Liver Transpl. · Jul 2008

    Comparative Study Clinical Trial

    Comparison of the USCOM ultrasound cardiac output monitor with pulmonary artery catheter thermodilution in patients undergoing liver transplantation.

    • Lai-Sze Grace Wong, Boon-Hun Yong, Karl Kang Young, Lee-Sung Lau, King-Lik Cheng, Jeff Siu-Fan Man, and Michael G Irwin.
    • Department of Anaesthesiology, Queen Mary Hospital, Hong Kong.
    • Liver Transpl. 2008 Jul 1;14(7):1038-43.

    AbstractThe aim of the study was to compare the standard technique of cardiac output determination by pulmonary artery catheter thermodilution (PAC-TD) with a noninvasive ultrasound Doppler monitor (USCOM Pty., Ltd., Coffs Harbour, Australia) in surgery for liver transplantation. We wished to determine if the degree of accuracy would allow the ultrasound cardiac output monitor (USCOM) to be used as an alternative monitor in a clinical setting in which wide fluctuations in cardiac output could be expected. This was a prospective method comparison study, with 71 paired measurements obtained in 12 patients undergoing liver transplantation in a university teaching hospital. Bland-Altman analysis of the 2 techniques showed a bias of 0.39 L/minute, with the USCOM cardiac output lower compared with that of PAC-TD. The bias was small and did not vary with the magnitude of the cardiac output. The 95% limits of agreement were -1.47 and 2.25 L/minute. There was good repeatability for USCOM measurements, with a repeatability coefficient of 0.43 for USCOM versus 0.77 for PAC-TD. We conclude that USCOM is acceptable for the clinical determination of noninvasive cardiac output, particularly in situations in which tracking changes over time is more important than knowing the precise value. However, the utility of USCOM is limited by its inability to measure pulmonary artery pressure.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…