• Perfusion · Jan 2006

    Comparative Study Clinical Trial

    Reliability of point-of-care hematocrit, blood gas, electrolyte, lactate and glucose measurement during cardiopulmonary bypass.

    • J Steinfelder-Visscher, P W Weerwind, S Teerenstra, and M H J Brouwer.
    • Department of Extra-Corporeal Circulation, Radboud University Nijmegen Medical Centre, Nijmegen, The Netherlands.
    • Perfusion. 2006 Jan 1;21(1):33-7.

    BackgroundRecently, the GEM Premier blood gas analyser was upgraded to the GEM Premier 3000. In addition to pH, pCO2, pO2, Na+, K+, Ca2+, and hematocrit measurement, glucose and lactate can be measured on the GEM Premier 3000. In this prospective clinical study, the analytical performance of the GEM Premier 3000 was compared with the Ciba Corning 865 analyser for blood gas/electrolytes/metabolites, and for hematocrit with the Sysmex XE 2100 instrument.MethodsDuring a 6-month period, 127 blood samples were analysed on both the GEM Premier 3000 analyser and our laboratory analysers (Ciba Corning 865/Sysmex 2100 instrument), and compared using the agreement analysis for quantitative data.ResultsWith the exception of K+, the other parameters (pCO2, pO2, Na+, Ca2+, hematocrit, glucose, and lactate) can be described in terms of the mean and standard deviation of the differences. For K+ measurement, a clear linear trend (r=0.79, p<0.001) in the deviation of the GEM Premier 3000 from the Ciba Corning was noticed, ie, in the lower or upper K+ reference range, the GEM Premier 3000 measured systematically too low or too high, respectively. Furthermore, in comparison with the other parameters, a therapeutically unacceptable systematic difference (mean of difference: -2.2%, p=0.05) in hematocrit measurement on the GEM Premier 3000 was observed for hematocrit values below 30%. The variance of the readings for the GEM Premier 3000 measurements was at clinically acceptable levels.ConclusionThe GEM Premier 3000 analyser seems to be suitable for point-of-care testing of electrolytes, metabolites, and blood gases during cardiopulmonary bypass. However, its downward bias in hematocrit values below 30% suggests that using the GEM Premier 3000 as a transfusion trigger leads to overtreatment with packed red cells.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.