• Pediatr Crit Care Me · May 2013

    Comparative Study

    Comparison of nasal continuous positive airway pressure delivered by seven ventilators using simulated neonatal breathing.

    • Thomas Drevhammar, Kjell Nilsson, Henrik Zetterström, and Baldvin Jonsson.
    • Department of Woman and Child Health, Karolinska Institutet, Stockholm, Sweden. thomas.drevhammar@jll.se
    • Pediatr Crit Care Me. 2013 May 1;14(4):e196-201.

    ObjectivesNasal continuous positive airway pressure (NCPAP) is an established treatment for respiratory distress in neonates. Most modern ventilators are able to provide NCPAP. There have been no large studies examining the properties of NCPAP delivered by ventilators. The aim of this study was to compare pressure stability and imposed work of breathing (iWOB) for NCPAP delivered by ventilators using simulated neonatal breathing.DesignExperimental in vitro study.SettingResearch laboratory in Sweden.InterventionNone.Measurements And Main ResultsNeonatal breathing was simulated using a mechanical lung simulator. Seven ventilators were tested at different CPAP levels using two breath profiles. Pressure stability and iWOB were determined. Results from three ventilators revealed that they provided a slight pressure support. For these ventilators, iWOB could not be calculated. There were large differences in pressure stability and iWOB between the tested ventilators. For simulations using the 3.4-kg breath profile, the pressure swings around the mean pressure were more than five times greater, and iWOB more than four times higher, for the system with the highest measured values compared with the system with the lowest. Overall, the Fabian ventilator was the most pressure stable system. Evita XL and SERVO-i were found more pressure stable than Fabian in some simulations. The results for iWOB were in accordance with pressure stability for systems that allowed determination of this variable.ConclusionsSome of the tested ventilators unexpectedly provided a minor degree of pressure support. In terms of pressure stability, we have not found any advantages of ventilators as a group compared with Bubble CPAP, Neopuff, and variable flow generators that were tested in our previous study. The variation between individual systems is great within both categories. The clinical importance of these findings needs further investigation.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.