- Jaehoon Oh, Youngjoon Chee, Taeho Lim, Boseung Kang, and Youngsoo Suh.
- Dept. of Emergency Medicine, College of Medicine, Hanyang University, Seoul, Republic of Korea.
- Resuscitation. 2012 Apr 1;83(4):500-4.
BackgroundChest compression (CC) feedback devices are used to perform CC measurements effectively and accurately on patients in hospital beds. However, these devices do not take account of the compression of the mattress, which results in overestimation of CC depth. In this study, we propose a new method using two accelerometers to overcome this limitation and thus measure compression depth more accurately when performing cardiopulmonary resuscitation (CPR) on patients.MethodOne accelerometer was placed on the manikin's sternum (a1), and the other between the manikin's back and the mattress (a2). The compression depth was calculated by integrating the acceleration twice using a digital signal processing technique. We compared CC depth from dual accelerometers and single accelerometer (a1) on the foam and inflated air mattress with eight CPR providers.ResultWhen CC was done on a manikin lying on the floor, there was no significant difference between measurement techniques (p>0.05). When CC was done on a manikin lying on the foam and inflated air mattress supporting system, our method significantly improved the estimation of CC depth, irrespective of the presence or absence of a backboard (p<0.001).ConclusionMeasuring CC depth using two accelerometers is more effective than using one in increasing the accuracy of CC depth estimation when CPR is performed on the foam and inflated air mattress, regardless of the presence or absence of a backboard.Copyright © 2011 Elsevier Ireland Ltd. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.