• Critical care medicine · Dec 1998

    Noninvasive cardiac output monitoring by aortic blood flow determination: evaluation of the Sometec Dynemo-3000 system.

    • A Cariou, M Monchi, L M Joly, F Bellenfant, Y E Claessens, D Thébert, F Brunet, and J F Dhainaut.
    • Medical Intensive Care Unit, Cochin-Port Royal University Hospital, Paris, France. alain.cariou@cch.ap-hop-paris.fr
    • Crit. Care Med. 1998 Dec 1;26(12):2066-72.

    ObjectiveThe Sometec Dynemo-3000 system allows the permanent measurement of descending aorta diameter by an echographic (A-scan) device and the blood flow velocity by a pulse Doppler velocimeter. The Dynemo-3000 then furnishes a new hemodynamic parameter, i.e., descending aortic blood flow (ABF), which is a fraction of the cardiac output (CO). We evaluate the ability of this system to measure the aortic diameter and to accurately detect ABF changes.DesignA case study prospective trial.SettingA 24-bed medical intensive care unit of a 1,100-bed university hospital.PatientsTwenty critically ill patients fully sedated, mechanically ventilated, and monitored by a pulmonary artery catheter.InterventionsCO values determined by conventional thermodilution method (TD-CO) and ABF were recorded during the study, which included two initial baseline periods, a dobutamine infusion (5 microg/kg/min) interval of 30 mins, and a third baseline period. To assess the accuracy of A-scan, aortic diameter was measured by transesophageal echocardiography. The difference between echocardiography and A-scan was used to determine bias and precision for aortic diameter measurements. TD-CO and ABF variations were analyzed using Kruskal-Wallis and Wilcoxon tests. Association between TD-CO and ABF values was determined by calculating the linear correlation coefficient. The ability of ABF to detect a TD-CO >6.0 L/min and its variations >13% was analyzed by determination of sensitivity, specificity, and positive (PPV) and negative (NPV) predictive values.Measurements And Main ResultsAortic diameter measurements by A-scan and bidimensional methods were 23.0+/-2.8 mm (SD) and 24.2+/-2.7 mm, respectively. Bias and precision were 1.1 mm and 1.4 mm (95% confidence interval: -1.9 to 3.7), respectively. During the course of dobutamine infusion, we observed a significant increase of TD-CO mean value from 6.65+/-1.53 L/min to 9.30+/-2.5 L/min (p=.0008), and a parallel and significant increase in ABF mean value from 4.34+/-1.18 L/min to 5.70+/-1.63 L/min (p= .0029). Absolute TD-CO and ABF values had a correlation coefficient of 0.80. For detection of an increased TD-CO, PPV and NPV were 87% and 86%, respectively. For detection of TD-CO changes >13%, PPV and NPV were 80% and 94%, respectively.ConclusionsThe Dynemo-3000 system is able to display the real aortic diameter, which is one of the most important components of this noninvasive ultrasonic technique. When compared with TD-CO, the ABF determination provided by this ultrasonic device constitutes a reliable noninvasive tool for estimating CO and tracking its changes.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…