- D S Gruszka, K J Burkhart, T E Nowak, T Achenbach, P M Rommens, and L P Müller.
- Department of Trauma Surgery, Centre for Musculoskeletal Surgery, The University Medical Center Mainz, Mainz, Germany. dominik.gruszka@unimedizin-mainz.de
- J Hand Surg Am. 2012 Jun 1;37(6):1142-50.
PurposeTo test a new generation of compression screws: the Acumed Acutrak 2 Mini (AA; Acumed, Hillsboro, OR), the Stryker TwinFix (ST; Stryker, Kalamazoo, MI), and the Synthes 3.0 headless compression screw (SH; Synthes, Solothurn, Switzerland).MethodsWe used 40 fresh-frozen human scaphoids for this study. Bone density was measured. A K-wire was inserted centrally. A perpendicular osteotomy was created in the middle third (Herbert B2 fracture). A custom-made load sensor was placed between the bone fragments. All screws were implanted according to the manufacturers' instructions. The Synthes 2.0 cortical screw (SC), implanted as a lag screw, was used as a reference. The compression force during each experiment was digitally monitored for 12 hours while the data were acquired. The data were analyzed using analysis of variance with the Bonferroni correction.ResultsImmediately after screw insertion, ST reached 226 N, followed by AA with 191 N, SH with 137 N, and SC with 72 N. After 12 hours, ST displayed the highest residual compression force, 141 N, followed by AA with 121 N, SH with 78 N, and SC with 32 N. The differences were significant for ST and AA compared to SC. The loss of compression force over 12 hours was 39% for ST, 42% for AA, 49% for SH, and 55% for SC.ConclusionsThe new generation of headless compression screws, especially ST and AA, provided significantly higher compression forces after 12 hours, as well as the least loss of compression force over time, in comparison to a classic cortical lag screw.Clinical RelevanceA new generation of headless compression screws, by producing higher compression forces, increase stability at the fracture site and might thereby promote bone healing.Copyright © 2012 American Society for Surgery of the Hand. Published by Elsevier Inc. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.