Pediatr Crit Care Me · Jan 2010
Cardiac output measurement using an ultrasound dilution method: a validation study in ventilated piglets.
- Willem P de Boode, Arno F J van Heijst, Jeroen C W Hopman, Ronald B Tanke, Hans G van der Hoeven, and Kian D Liem.
- Department of Neonatology, Radboud University Nijmegen Medical Centre, Nijmegen, Netherlands. W.deBoode@cukz.umcn.nl
- Pediatr Crit Care Me. 2010 Jan 1;11(1):103-8.
ObjectiveTo assess agreement between a new method of cardiac output monitoring, using ultrasound dilution technology and ultrasound transit time-based measurement of pulmonary blood flow in a piglet model.DesignProspective, experimental juvenile animal study.SettingAnimal laboratory of a university hospital.SubjectsNine random-bred piglets.InterventionsAfter the animals received general anesthesia, we placed intravascular arterial and central venous catheters with the tip positioned in the abdominal aorta and the right atrium, respectively. The catheters were connected to the ultrasound dilution cardiac output monitor. An ultrasound transit time perivascular flow probe was positioned around the common pulmonary artery and served as the standard reference measurement. Cardiac output was manipulated during the experiment by creating hemorrhagic hypotension. Ultrasound dilution cardiac output was measured intermittently with injection volumes of 0.5 mL/kg and 1.0 mL/kg of isotonic saline at body temperature.Measurements And Main ResultsUltrasound dilution cardiac output (Q) measurement was compared with pulmonary blood flow (Q). Bias, defined as Q minus Q, was calculated for each measurement. Mean bias with standard deviation was calculated for measurements with volumes of injected saline, 0.5 mL/kg and 1.0 mL/kg, and compared using the Mann-Whitney U test. Mean bias (sd) between Q and Q was 0.040 (0.132) and 0.058 (0.136) L/min for measurement with 0.5 mL/kg and 1.0 mL/kg of isotonic saline, respectively (no statistically significant difference).ConclusionsUltrasound dilution cardiac output measurement is reliable in piglets with the use of a small volume of a nontoxic indicator (isotonic saline).
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.