• J Cardiovasc Magn Reson · Jan 2014

    Immediate and 12 months follow up of function and lead integrity after cranial MRI in 356 patients with conventional cardiac pacemakers.

    • Olaf M Muehling, Reza Wakili, Martin Greif, Franz von Ziegler, Dominik Morhard, Hartmut Brueckmann, and Alexander Becker.
    • Cardiology Clinic Harlaching, University of Munich, Munich, Germany. DrMuehling@aol.com.
    • J Cardiovasc Magn Reson. 2014 Jan 1;16:39.

    BackgroundConventional cardiac pacemakers are still often regarded as a contraindication to magnetic resonance imaging (MRI). We conducted this study to support the hypothesis that it is safe to scan patients with cardiac pacemakers in a 1.5 Tesla MRI, if close supervision and monitoring as well as adequate pre- and postscan programming is provided.MethodsWe followed up 356 patients (age 61.3 ± 9.1 yrs., 229 men) with single (n = 132) or dual chamber (n = 224) cardiac pacemakers and urgent indication for a cranial MRI for 12 months. The scans were performed at 1.5T. During the scan patients were monitored with a 3-lead ECG and pulse oximetry. Prior to the scan pacemakers were programmed according to our own protocol.ResultsAll 356 scans were completed without complications. No arrhythmias were induced, programmed parameters remained unchanged. No pacemaker dysfunction was identified. Follow-up examinations were performed immediately, 2 weeks, 2, 6, and 12 months after the scan. There was no significant change of pacing capture threshold (ventricular 0.9 ± 0.4 V@0.4 ms, atrial 0.9 ± 0.3 V@0.4 ms) immediately (ventricular 1.0 ± 0.3 V@0.4 ms, atrial 0.9 ± 0.4 V@0.4 ms) or at 12 months follow-up examinations (ventricular 0.9 ± 0.2 V@0.4 ms, atrial 0.9 ± 0.3 V@0.4 ms). There was no significant change in sensing threshold (8.0 ± 4.0 mV vs. 8.1 ± 4.2 mV ventricular lead, 2.0 ± 0.9 mV vs. 2.1 ± 1.0 mV atrial lead) or lead impedance (ventricular 584 ± 179 Ω vs. 578 ± 188 Ω, atrial 534 ± 176 Ω vs. 532 ± 169 Ω) after 12 months.ConclusionsThis supports the evidence that patients with conventional pacemakers can safely undergo cranial MRI in a 1.5T system with suitable preparation, supervision and precautions. Long term follow-up did not reveal significant changes in pacing capture nor sensing threshold.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.