Anaesth Intensive Care · Oct 1993
The Australian Incident Monitoring Study. The pulse oximeter: applications and limitations--an analysis of 2000 incident reports.
- W B Runciman, R K Webb, L Barker, and M Currie.
- Department of Anaesthesia and Intensive Care, University of Adelaide, S.A.
- Anaesth Intensive Care. 1993 Oct 1;21(5):543-50.
AbstractThe first 2000 incidents reported to the Australian Incident Monitoring Study were analysed with respect to the role of the pulse oximeter. Of these 184 (9%) were first detected by a pulse oximeter and there were a further 177 (9%) in which desaturation was recorded. Of the 1256 incidents which occurred in association with general anaesthesia 48% were "human detected" and 52% "monitor detected". The pulse oximeter was ranked first and detected 27% of these monitor detected incidents; this figure would have been over 40% if an oximeter had always been used and its more informative modulated pulse tone relied upon instead of that of the "bleep" of the ECG. The pulse oximeter is the "front-line" monitor for endobronchial intubation, the fourth most common incident in association with general anaesthesia (it detected 87% of the 76 cases in which it was in use). It also played an invaluable role as a "back-up" monitor in 40 life-threatening situations in which "front-line" monitors (e.g. oxygen analyser, low pressure alarm, capnograph) were either not in use, were being used incorrectly or failed. Other situations detected, in order of frequency of detection, were: circuit disconnection, circuit leak, desaturation (severe shunt), oesophageal intubation, aspiration and/or regurgitation, pulmonary oedema, endotracheal tube obstruction, severe hypotension, failure of oxygen delivery, hypoxic gas mixture, hypoventilation, anaphylaxis, air embolism, bronchospasm, malignant hyperthermia, and tension pneumothorax.(ABSTRACT TRUNCATED AT 250 WORDS)
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.