Comparison of the accuracy of noninvasive hemoglobin monitoring by spectrophotometry (SpHb) and HemoCue® with automated laboratory hemoglobin measurement.
- Lionel Lamhaut, Roxana Apriotesei, Xavier Combes, Marc Lejay, Pierre Carli, and Benoît Vivien.
- Department of Anesthesiology and Intensive Care, Hôpital Necker-Enfants Malades, Assistance Publique Hôpitaux de Paris, and Université Paris Descartes, Paris, France.
- Anesthesiology. 2011 Sep 1;115(3):548-54.
BackgroundThe reference method for hemoglobin concentration measurement remains automated analysis in the laboratory. Although point-of-care devices such as the HemoCue® 201+ (HemoCue, Ängelholm, Sweden) provide immediate hemoglobin values, a noninvasive, spectrophotometry-based technology (Radical-7®; Masimo Corp., Irvine, CA) that provides continuous online hemoglobin (SpHb) measurements has been introduced. This clinical study aimed to test the hypothesis that SpHb monitoring was equivalent to that of HemoCue® (the automated hemoglobin measurement in the laboratory taken as a reference method) during acute surgical hemorrhage.MethodsBlood for laboratory analysis was sampled after induction of anesthesia, during surgery according to the requirements of the anesthesiologist, and finally after the transfer of the patient to the recovery room. When each blood sample was taken, capillary samples were obtained for analysis with HemoCue®. SpHb monitoring was performed continuously during surgery. Using the automated hemoglobin measurement in the laboratory as a reference method, the authors tested the hypothesis that SpHb monitoring is equivalent to that of HemoCue®. The agreement between two methods was evaluated by linear regression and Bland and Altman analysis.ResultsEighty-five simultaneous measurements from SpHb, HemoCue®, and the laboratory were obtained from 44 patients. Bland and Altman comparison of SpHb and HemoCue® with the laboratory measurement showed, respectively, bias of -0.02 ± 1.39 g · dl(-1) and -0.17 ± 1.05 g · dl(-1), and a precision of 1.11 ± 0.83 g · dl(-1) and 0.67 ± 0.83 g · dl(-1). Considering an acceptable difference of ± 1.0 g · dl(-1) with the laboratory measurement, the percentage of outliers was significantly higher for SpHb than for HemoCue® (46% vs. 16%, P < 0.05).ConclusionsTaking automated laboratory hemoglobin measurement as a reference, the study shows that SpHb monitoring with Radical-7® gives lower readings than does the HemoCue® for assessment of hemoglobin concentration during hemorrhagic surgery.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.