• Pacing Clin Electrophysiol · Apr 2000

    Comparative Study

    Efficacy and safety of bipolar sensing with high atrial sensitivity in dual chamber pacemakers.

    • U K Wiegand, F Bode, W Peters, H Haase, H Bonnemeier, H A Katus, and J Potratz.
    • Medical University of Luebeck, Department of Internal Medicine II, Germany.
    • Pacing Clin Electrophysiol. 2000 Apr 1;23(4 Pt 1):427-33.

    AbstractIn dual chamber pacemakers, atrial sensing performance is decisive for maintenance of AV synchrony. Particularly, the efficacy of mode switching algorithms during intermittent atrial tachyarrhythmias depends on the sensitive detection of low potential amplitudes. Therefore, a high atrial sensitivity of 0.18 mV, commonly used in single lead VDD pacemakers, was investigated for its efficacy and safety in DDD pacing. Thirty patients received dual chamber pacemakers and bipolar atrial screw-in leads for sinus node syndrome or AV block; 15 patients suffered from intermittent atrial fibrillation. Pace makers were programmed to an atrial sensitivity of 0.18 mV. Two weeks, 3, 9, and 15 months after implantation, P wave sensing threshold and T wave oversensing thresholds for the native and paced T wave were determined. The myopotential oversensing thresholds were evaluated by isometric contraction of the pectoral muscles. Automatic mode switch to DDIR pacing was activated when the mean atrial rate exceeded 180 beats/min. The patients were followed by 24-hour Holter monitoring. Two weeks after implantation, mean atrial sensing threshold was 1.81 +/- 0.85 mV (range 0.25-2.8 mV) without significant differences during further follow-up. Native T wave sensing threshold was < 0.18 mV in all patients. In 13% of patients, paced T waves were perceived in the atrial channel at the highest sensitivity. This T wave sensing could easily be avoided by programming a postventricular atrial refractory period exceeding 300 ms. Myopotential oversensing could not be provoked and Holter records showed no signs of sensing dysfunction. During a 15-month follow-up, 1,191 mode switch events were counted by autodiagnostic pacemaker function. Forty-two of these events occurred during Holter monitoring. Unjustified mode switch was not observed. In DDD pacemakers, bipolar atrial sensing with a very high sensitivity is efficient and safe. Using these sensitivity settings, activation of the mode switch algorithm almost completely avoids fast transmission of atrial rate to the ventricle during atrial fibrillation.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…