    Reliability and validity of a new fingertip-shaped pressure algometer for assessing pressure pain thresholds in the temporomandibular joint and masticatory muscles.

    • Olaf Bernhardt, Eric L Schiffman, and John O Look.
    • Department of Restorative Dentistry, School of Dentistry, University of Greifswald, Greifswald, Germany. obernhar@uni-greifswald.de
    • J Orofac Pain. 2007 Jan 1;21(1):29-38.

    AimsTo test in vitro and in vivo the reliability and accuracy of a new algometer, the pressure algometer for palpation (PAP), for measuring pressure pain thresholds (PPTs) and to compare its features with those of a commercially available pressure algometer.MethodsFor in vitro accuracy testing, 6 repeated measurements were made at 8 defined test weights from 0.5 to 5 lb. In vivo validity testing compared the PAP to a standard instrument, the hand-held Somedic algometer, at 16 sites including the masticatory muscles, the temporomandibular joints, and the frontalis (as the control site) in 15 temporomandibular disorder (TMD) cases and 15 controls. Intraexaminer reliability was also assessed for both algometers.ResultsIn vitro reliability was high, with coefficients of variation of < 5% and a single-measurement standard deviation of 2.1 kPa. Accuracy was also high, with PAP measurements correlating with test weights at r = .99 (P < .001). Repeated measures reliability in vivo was high, with intraclass correlation estimates of 0.73 to 0.96 for the PAP and 0.78 to 0.99 for the Somedic algometer. PPT values correlated moderately between the 2 devices (r ranged from 0.38 to 0.66; P < or = .05) and were consistently higher for the PAP at all sites (P < .001). Differences between controls and TMD cases were also significant for both algometers (P < .006).ConclusionBoth the PAP and the Somedic algometer showed high reliability. Concurrent validity was demonstrated by statistically significant correlations between the devices. Both showed equally high capacity for differentiating TMD cases from controls. The PAP yielded significantly higher PPTs than the Somedic algometer.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.