Emerg Med Australas · Aug 2009
Validation of a point of care troponin assay in real life Emergency Department conditions.
- Conrad Loten, John Attia, Carolyn Hullick, John Marley, and Patrick McElduff.
- John Hunter Hospital, Division of Emergency Medicine, University of Newcastle, New South Wales, Australia. conradloten@gmail.com
- Emerg Med Australas. 2009 Aug 1;21(4):286-92.
ObjectiveTo validate the accuracy of a Point of care (POC) troponin device (Abbott i-Stat) in real life ED conditions.MethodsDesignA three-way comparison between troponin I results obtained by experienced POC operators, inexperienced ED staff and central laboratory criterion standard.ParticipantsConvenience sample of 332 patients presenting to the ED with possible coronary syndromes.ResultsSpearman correlation coefficient for experienced versus laboratory was 0.83 (95% CI 0.78-0.87), occasional users versus laboratory was 0.76 (95% CI 0.71-0.81), and experienced versus occasional users on POC was 0.82 (95% CI 0.76-0.87). Using local troponin cut-off of 0.1 ng/mL, kappa coefficient was 0.94 for occasional users versus laboratory, 0.91 for experienced versus laboratory and 0.94 for experienced versus occasional users. Bland-Altman plots showed good agreement across the range of measured values. The sensitivity of i-Stat (vs laboratory as criterion standard) was 92.2% (95% CI 83.8%-97.0%) with the local cut-off but only 70.1% (95% CI 60.5%-78.6%) using the lowest cut-off associated with acceptable reproducibility (10% coefficient of variation).ConclusionsThe i-Stat POC device produces similar results in the hands of experienced and occasional operators in ED. There is good agreement between the POC and laboratory at levels used to diagnose infarction by older, more specific criteria. When compared using new lower cut-offs, the i-Stat had poor sensitivity.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.