    Effects of different triggering systems and external PEEP on trigger capability of the ventilator.

    • Y A Konyukov, N Kuwayama, T Fukuoka, T Takahashi, T Mayumi, T Hotta, and J Takezawa.
    • Department of Emergency and Intensive Care Medicine, Nagoya University School of Medicine, Japan.
    • Intensive Care Med. 1996 Apr 1;22(4):363-8.

    ObjectiveThe triggering capability of both the pressure and flow triggering systems of the Servo 300 ventilator (Siemens-Elema, Sweden) was compared at various levels of positive end-expiratory pressure (PEEP), airway resistance (R(aw)), inspiratory effort and air leak, using a mechanical lung model.DesignThe ventilator was connected to a two bellows-in-series-type lung model with various mechanical properties. Lung compliance and chest wall compliance were 0.03 and 0.121/cmH2O, respectively. R(aw) was 5, 20 and 50 cmH2O/l/s. Respiratory rate was 15 breaths/min. To compare the triggering capability of both systems, the sensitivity of pressure and flow triggered pressure support ventilation (PSV) was adjusted to be equal by observing the triggering time at 0 cmH2O PEEP and 16 cmH2O of pressure support (PS) with no air leak. No auto-PEEP was developed. In the measurement of trigger delay, the PS level ranged from 16 to 22 cmH2O to attain a set tidal volume (V(T)) of 470 ml at a R(aw) of 5, 20 and 50 cmH2O/l/s. The PEEP level was then changed from 0, 5 and 10 cmH2O at a PS level of 17 cmH2O and R(aw) of 5 and 20 cmH2O/l/s, and the trigger delay was determined. The effect of various levels of air leak and inspiratory effort on triggering capability was also evaluated. Inspiratory effort during triggering delay was estimated by measurements of pressure differentials of airway pressure (Paw) and driving pressure in the diaphragm bellows (Pdriv) in both systems.Measurements And ResultsThere were no significant differences in trigger delay between the two triggering systems at the various PEEP and R(aw) levels. At the matched sensitivity level, air leak decreased trigger delay in both systems, and additional PEEP caused auto-cycling. A low inspiratory drive increased trigger delay in the pressure sensing system, while trigger delay was not affected in the flow sensing system. The Paw and Pdriv differentials were lower in flow triggering than in pressure triggering.ConclusionsWith respect to triggering delay, the triggering capabilities of the pressure and flow sensing systems were comparable with and without PEEP and/or high airway resistance at the same sensitivity level, unless low inspiratory drive and air leak were present. In terms of pressure differentials, the flow triggering system may require less inspiratory effort to trigger the ventilator than that of the pressure triggering system with a comparable triggering time. However, this difference may be extremely small.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…