Comparative Study
Clinical evaluation of pushover mechanical ventilation with the Ohmeda Universal Portable Anesthesia Complete vaporizer.
- J K Hawkins, S A Ciresi, and W J Phillips.
- Anesthesia and Operative Nursing Services, Womack Army Medical Center, Fort Bragg, NC 28307, USA.
- Mil Med. 1998 Mar 1;163(3):164-8.
AbstractPrevious studies have not demonstrated the usefulness of a mechanical ventilator with the Universal Portable Anesthesia Complete (UPAC) field anesthesia delivery system in a pushover mode. This study demonstrated that the Lifecare PLV-100 ventilator can function effectively in a practical pushover configuration with the UPAC vaporizer. By comparison, vaporizer output followed the patterns of documented concentration curves for isoflurane at a given dial setting and minute ventilation. Measured airway pressures in the breathing circuit were within physiological parameters.