• Intensive care medicine · Mar 1996

    Comparative Study Clinical Trial

    Accuracy of oscillometric blood pressure measurement in critically ill neonates with reference to the arterial pressure wave shape.

    • M Gevers, H R van Genderingen, H N Lafeber, and W W Hack.
    • Department of Pediatrics-neonatology, Free University Hospital, Amsterdam, The Netherlands.
    • Intensive Care Med. 1996 Mar 1;22(3):242-8.

    ObjectiveTo perform further evaluation of the oscillometric device for neonatal arterial blood pressure (ABP) measurement, using a catheter-manometer system (CMS) for accurate intraarterial measurement. We aimed to describe the influence of the radial artery wave shape on oscillometric ABP determination, as pressure wave-shape influences the relationships between systolic arterial pressure (SAP), diastolic arterial pressure (DAP) and mean arterial pressure (MAP) in the wave. These relationships are part of the algorithms contributing to the final ABP determination in the oscillometric device.DesignIntra-patient comparison of two blood pressure measurement systems.SettingNeonatal intensive care unit.PatientsIn 51 critically ill newborn infants, ABP was determined oscillometrically in the brachial artery and, simultaneously, invasively in the radial artery using a high-fidelity CMS. Clinical data of the infants were: gestational age: 29 (25-41) weeks; birthweight: 1200 (500-3675) g, postnatal age: 6 (2-46) h.MethodsStatistical analysis was performed with the paired Student's t-test. Multiple regression analysis was used to determine the influence of birthweight and height of the blood pressure on the results.Measurements And Main ResultsIn 51 infants, 255 paired values of SAP, DAP and MAP were recorded. In all recordings, we determined the relationship between SAP, DAP and MAP, using the equation: MAP = alpha%(SAP - DAP) + DAP. For SAP, DAP, MAP and alpha, we computed mean differences (bias) and the limits of agreement (precision). Biases for SAP, DAP, MAP and alpha were significantly different from zero (P < 0.001) and the limits of agreement for SAP, DAP and MAP were wide: 18.8 mmHg, 17.2 mmHg and 15.2 mmHg respectively. The relationship between invasive and noninvasive values is only partly (7-19%) influenced by the height of the blood pressure; low values of SAP, DAP and MAP tend to give overestimated oscillometric values. In the relationship between SAP, DAP and MAP, alpha was found to be 47% invasively (as generally found in the radial artery in newborns) and 34% noninvasively (as generally found in the brachial/radial artery in adults).ConclusionsInaccuracy of the oscillometric device may be partly explained by the incorporation of an inappropriately fixed algorithm for final ABP determination in newborns. Care should be taken when interpreting the oscillometrically derived values in critically ill newborn infants.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…