Comparative Study Clinical TrialMeasurement of mean arterial pressure: comparison of the Vasotrac monitor with the finger differential oscillometric device.
- K Jagomägi, R Raamat, J Talts, U Ragun, and P Tähepõld.
- Institute of Physiology, University of Tartu, and Department of Anaesthesiology and Intensive Care, Tartu University Hospital, Tartu, Estonia. kersti.jagomagi@ut.ee
- Physiol Res. 2010 Jan 1;59(5):691-6.
AbstractThe Vasotrac monitor provides non-invasive near-continuous blood pressure monitoring and is designed to be an alternative to direct intra-arterial blood pressure (BP) measurement. As compared to radial artery invasive BP and upper arm non-invasive BP, Vasotrac readings have been found to have a good agreement with them. However, discrepancies have been reported when rapid changes in BP exist. In the present study we compared BP measured by the Vasotrac monitor on the radial artery with that recorded on the finger arteries by the differential oscillometric device allowing measurement on the beat-to-beat basis. Comparisons were performed on the mean arterial pressure (MAP) level. Special attention was paid to the signal conditioning before comparison of pressures of different temporal resolution. Altogether 383 paired MAP measurements were made in 14 healthy subjects. Based on all 383 paired measurements, the MAP values measured at the radial artery at rest were 4.8+/-6.0 mm Hg higher than those measured on fingers. The observed difference between the Vasotrac and differential oscillometric device can be explained by different measurement sites. This result is consistent with previous investigations, and the Vasotrac monitor can be considered to adequately track relatively rapid MAP changes on the radial artery. Attention should be paid to a proper signal conditioning before comparison of results obtained by different devices.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.