    Misplacement of a psoas compartment catheter in the subarachnoid space.

    • Rainer J Litz, Oliver Vicent, Diana Wiessner, and Axel R Heller.
    • Department of Anesthesiology and Intensive Care Medicine, University Hospital Carl Gustav Carus, Dresden, Germany. rainer.litz@mailbox.tu-dresden.de
    • Reg Anesth Pain Med. 2004 Jan 1; 29 (1): 60-4.

    Background And ObjectivesThis case report describes an unusual cause of misplacement of an indwelling catheter in the subarachnoid space after primary psoas compartment block in a patient undergoing total knee arthroplasty.Case ReportA 67-year-old woman presenting for total knee joint replacement received a combination of continuous psoas compartment block and sciatic nerve block. Neurostimulation and additional ultrasound guidance were used for identification of the lumbar plexus. After elicitation of a quadriceps motor response, a negative aspiration test, and an uneventful test dose, 20 mL ropivacaine 0.375% and 20 mL mepivacaine 1% were injected. Despite difficult ultrasound conditions because of intestinal air, local anesthetic spread was observed paravertebrally at the medial border of the psoas muscle as usual. A catheter was then advanced 7 cm through the insulated directional puncture needle. An additional sciatic nerve block was performed by using Labat's approach. Ten minutes after injection unilateral sensory block was noted and surgery was started. After uneventful surgery, bilateral sensory block to the T4 level and complete motor block in both lower limbs was detected. A second aspiration test was negative, and an epidural block was suspected. For verification of the catheter tip location, a computed tomography scan with contrast dye was performed revealing catheter placement in the subarachnoid space. The catheter was removed and showed a kink about 7 cm from the tip. After regression of the neuraxial block, lumbar plexus block persisted for another 2 hours.ConclusionAn additional test dose via the catheter is recommended if the indwelling catheter is inserted after injection of the local anesthetics through the puncture needle. If epidural anesthesia occurs, an x-ray of the catheter is advisable because negative aspiration via catheter does not rule out subarachnoid catheter location.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…