    Errors in 14 pulse oximeters during profound hypoxia.

    • J W Severinghaus, K H Naifeh, and S O Koh.
    • Department of Anesthesia, University of California, San Francisco 94143-0542.
    • J Clin Monit. 1989 Apr 1;5(2):72-81.

    AbstractThe accuracy of pulse oximeters from fourteen manufacturers was tested during profound brief hypoxic plateaus in 125 subject sets using 50 normal adult volunteers, of whom 29 were studied two to nine times. A data set usually consisted of 10 subjects, and 13 sets were collected between August 1987 and July 1988. In the first 6 sets, six 30-second hypoxic plateaus were obtained per subject at 55 +/- 6% oxyhemoglobin (O2Hb) (range, 40 to 70%). In the last 7 sets, three hypoxic plateaus were obtained at each of four levels, approximately 86, 74, 62, and 50% O2Hb, for the purpose of linear regression analysis. Inspired oxygen was adjusted manually breath by breath in response to arterial oxygen saturation computed on-line from end-tidal oxygen and carbon dioxide tensions. End-plateau arterial blood O2Hb was analyzed by a Radiometer OSM-3 oximeter, and plateau pulse oximeter saturation (SpO2) was read by cursor from a computer record of the analog output. Three to 13 instruments were tested simultaneously by using 1 to 3 duplicate instruments from each of one to seven manufacturers. Variations introduced by manufacturers were tested on subsequent sets in several instruments. An index of error, "ambiguity" (alpha) of oxygen saturation, was defined as the absolute sum of bias and precision (mean and SD of SpO2 - O2Hb) at O2Hb = 55.8 +/- 4.5%, preserving the sign when bias was significant at P less than 0.05. Ambiguity values for finger probes (unless specified) with latest data were: Physio-Control, 3.9 (ear, 3.3); Puritan-Bennett, -4.4; Criticare, 5.8 (forehead, 4.7); Kontron, 5.9 (infant probe) and 6.1 (ear, 5.8; forehead, 7.1); Biochem, -6.0; Datex 6.4 (ear, 6.9; forehead, 6.8); Critikon, 8.4; SiMed, 8.6; Marquest, 9.0; Novametrix, 10.2; Invivo, -12.2 (ear, -14.3); Nellcor, -15.1; Ohmeda, -21.2; and Radiometer, -21.2 (ear, -9.6). Linear regression slopes of 36 instruments from twelve manufacturers generally deviated from 1 in proportion to alpha. The data showed substantial differences in bias and precision between pulse oximeters at low saturations, the most common problems being underestimation of saturation and failing precision.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.