- Health Devices. 2010 Dec 1;39(12):454-6.
AbstractDue to a software issue, the Moog Curlin 6000 CMS infusion pump may activate an error code associated with a hardware failure (error code 45) when no such failure exists. When this occurs, the unit will stop infusing, and the user may be unable to restart it. Moog has announced plans to release new software in the coming weeks to address this problem. Until the software is available, facilities whose units exhibit this error code should contact Moog for guidance.