Comparative StudyAccuracy of bedside glucose measurement from three glucometers in critically ill patients.
- Cornelia W E Hoedemaekers, Jacqueline M T Klein Gunnewiek, Marieke A Prinsen, Johannes L Willems, and Johannes G Van der Hoeven.
- Departments of Intensive Care, Radboud University Nijmegen Medical Centre, Nijmegen, The Netherlands. c.hoedemaekers@ic.umcn.nl
- Crit. Care Med. 2008 Nov 1;36(11):3062-6.
ObjectiveImplementation of strict glucose control in most intensive care units has resulted in increased use of point-of-care glucose devices in the intensive care unit. The aim of this study was to determine the reliability of point-of-care testing glucose meters among critically ill patients under intensive insulin treatment.DesignProspective observational study.PatientsIntensive care unit and non-intensive care unit patients in a tertiary care teaching hospital.MeasurementsA glucose oxidase method was used to validate the point-of-care testing devices. Three different point-of-care testing devices, Accu-Chek Sensor (Roche Diagnostics), Precision (Abbott Diagnostics), and HemoCue were tested. Glucose measurements were performed in duplicate by an experienced technician under standardized conditions in the hospital's laboratory, using arterial (intensive care unit patients) and arterial or venous (non-intensive care unit patients) heparinized whole blood samples.Main ResultsA strong correlation was found between the glucose oxidase method and the Accu-Chek device (r = .9596, p < 0.001). Mean absolute difference between the glucose oxidase and Accu-Chek was -0.32 mmol/L (95% confidence interval -0.84 to 1.48 mmol/L). Using the International Organization for Standardization (ISO) criteria, 27 of 197 samples (13.7%) were inaccurate. In all samples that failed to meet the ISO criteria, glucose values measured by the Accu-Chek device were higher compared with the glucose oxidase method. In another set of experiments among intensive care unit patients, strong positive correlations were also found between the other point-of-care testing devices and the glucose oxidase method. However, paired samples from Accu-Chek, HemoCue, and Precision failed the ISO criteria in 9 of 82 (11.0%), 4 of 82 (4.9%), and 11 of 82 (13.4%) of cases, respectively. In non-intensive care unit patients paired samples from Accu-Chek, HemoCue, and Precision failed the ISO criteria in 3 of 120 (2.5%), 11 of 120 (9.2%), and 16 of 120 (13.3%) cases, respectively.ConclusionsUnder standardized conditions, glucose results from three point-of-care testing devices were inaccurate in both intensive care unit and non-intensive care unit patients. Among intensive care unit patients, inaccurate glucose readings were most frequently falsely elevated, resulting in misinterpretation of high glucose values with subsequent inappropriate insulin administration or masking of true hypoglycemia.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:

- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.