• J Aerosol Med Pulm Drug Deliv · Jun 2015

    Comparative Study

    Influence of inspiratory flow pattern and nebulizer position on aerosol delivery with a vibrating-mesh nebulizer during invasive mechanical ventilation: an in vitro analysis.

    • Jonathan Dugernier, Xavier Wittebole, Jean Roeseler, Jean-Bernard Michotte, Thierry Sottiaux, Thierry Dugernier, Pierre-François Laterre, and Gregory Reychler.
    • 1Intensive Care Unit, University Hospital Saint-Luc, Brussels, Belgium.
    • J Aerosol Med Pulm Drug Deliv. 2015 Jun 1; 28 (3): 229-36.

    BackgroundAerosol delivery during invasive mechanical ventilation (IMV) depends on nebulizer type, placement of the nebulizer and ventilator settings. The purpose of this study was to determine the influence of two inspiratory flow patterns on amikacin delivery with a vibrating-mesh nebulizer placed at different positions on an adult lung model of IMV equipped with a proximal flow sensor (PFS).MethodsIMV was simulated using a ventilator connected to a lung model through an 8-mm inner-diameter endotracheal tube. The impact of a decelerating and a constant flow pattern on aerosol delivery was evaluated in volume-controlled mode (tidal volume 500 mL, 20 breaths/min, inspiratory time of 1 sec, bias flow of 10 L/min). An amikacin solution (250 mg/3 mL) was nebulized with Aeroneb Solo(®) placed at five positions on the ventilator circuit equipped with a PFS: connected to the endotracheal tube (A), to the Y-piece (B), placed at 15 cm (C) and 45 cm upstream of the Y-piece (D), and placed at 15 cm of the inspiratory outlet of the ventilator (E). The four last positions were also tested without PFS. Deposited doses of amikacin were measured using the gravimetric residual method.ResultsAmikacin delivery was significantly reduced with a decelerating inspiratory flow pattern compared to a constant flow (p<0.05). With a constant inspiratory flow pattern, connecting the nebulizer to the endotracheal tube enabled similar deposited doses than these obtained when connecting the nebulizer close to the ventilator. The PFS reduced deposited doses only when the nebulizer was connected to the Y-piece with both flow patterns or placed at 15 cm of the Y-piece with a constant inspiratory flow (p<0.01).ConclusionsUsing similar tidal volume and inspiratory time, a constant flow pattern (30 L/min) delivers a higher amount of amikacin through an endotracheal tube compared to a decelerating inspiratory flow pattern (peak inspiratory flow around 60 L/min). The optimal nebulizer position depends on the inspiratory flow pattern and the presence of a PFS.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.