Comparative StudyTransapical aortic valve replacement under real-time magnetic resonance imaging guidance: experimental results with balloon-expandable and self-expanding stents.
- Keith A Horvath, Dumitru Mazilu, Ozgur Kocaturk, and Ming Li.
- Cardiothoracic Surgery Research Program, National Heart, Lung and Blood Institute, National Institutes of Health, Bethesda, MD, USA. horvathka@mail.nih.gov
- Eur J Cardiothorac Surg. 2011 Jun 1;39(6):822-8.
ObjectiveAortic valves have been implanted on self-expanding (SE) and balloon-expandable (BE) stents minimally invasively. We have demonstrated the advantages of transapical aortic valve implantation (tAVI) under real-time magnetic resonance imaging (rtMRI) guidance. Whether there are different advantages to SE or BE stents is unknown. We report rtMRI-guided tAVI in a porcine model using both SE and BE stents, and compare the differences between the stents.MethodsA total of 22 Yucatan pigs (45-57 kg) underwent tAVI. Commercially available stentless bioprostheses (21-25 mm) were mounted on either BE platinum-iridium stents or SE-nitinol stents. rtMRI guidance was employed as the intraoperative imaging. Markers on both types of stents were used to enhance visualization in rtMRI. Pigs were allowed to survive and had follow-up MRI scans and echocardiography at 1, 3, and 6 months postoperatively.ResultsrtMRI provided excellent visualization of the aortic valve implantation mounted on both stent types. The implantation times were shorter with the SE stents (60 ± 14s) than with the BE stents (74 ± 18s), (p=0.027). The total procedure time was 31 and 37 min, respectively (p=0.12). It was considerably easier to manipulate the SE stent during deployment, without hemodynamic compromise. This was not always the case with the BE stent, and its placement occasionally resulted in coronary obstruction and death. Long-term results demonstrated stability of the implants with preservation of myocardial perfusion and function over time for both stents.ConclusionsSE stents were easier to position and deploy, thus leading to fewer complications during tAVI. Future optimization of SE stent design should improve clinical results.Copyright © 2010 European Association for Cardio-Thoracic Surgery. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.