Anesthesia and analgesia · Jul 2000
Case ReportsNitrogen purging of oxygen pipelines: an unusual cause of intraoperative hypoxia.
- A M Elizaga and R L Frerichs.
- Department of Anesthesiology, Auburn Regional Medical Center, Washington 98002-9906, USA. aelizaga@u.washington.edu
- Anesth. Analg. 2000 Jul 1;91(1):242-3.
AbstractIntraoperative hypoxia occurred in two patients during the maintenance of the medical gas system. Engineers were purging oxygen pipelines with nitrogen to remove particulate debris but were unaware of a connection to operating room pipelines. This case illustrates the importance of communication between anesthesia providers and engineers servicing the gas system.