Case Reports
A unique access for the ablation catheter to treat electrical storm in a patient with extracorporeal life support.
- Ekrem Ücer, Sabine Fredersdorf, Carsten Jungbauer, Kurt Debl, Alois Philipp, Matthias Amann, Andreas Holzamer, Andreas Keyser, Michael Hilker, Andreas Luchner, Christof Schmid, Günter Riegger, and Dierk Endemann.
- Department of Internal Medicine II Cardiology, University Hospital Regensburg, 93053 Regensburg, Germany.
- Europace. 2014 Feb 1;16(2):299-302.
AimsExtracorporeal membrane oxygenation (ECMO) is a very effective bridging therapy in patients with cardiogenic shock. To perform coronary angiography in these patients our group developed an unique system to get urgent vascular access with minimal additional vascular complication risk. The 6 Fr coronary catheters are introduced through a standard Y-connector, which is inserted into the arterial cannula of the ECMO-line close to the patient, the blind end of which is then equipped with a haemostatic valve (Check-Flo Performer accessory adapter, Cook Medical, USA). To the best of our knowledge, we here present the first patient, in whom this system had been used to insert an 8 Fr radiofrequency ablation catheter to treat incessant ventricular fibrillation.Methods And ResultsA 66-year-old patient had been transferred with electrical storm 5 days after an acute MI. After failed interventional and medical therapies an ECMO system had been inserted (right femoral artery cannula 15 Fr, left femoral vein cannula 21 Fr) and an electrophysiological study had been performed because of incessant ventricular fibrillation episodes, which always were induced by the same ventricular premature beat (VPB). During this first EP study over the left femoral artery the VPB could be targeted and successfully ablated. Unfortunately the VPB recovered again after some days so a second EP study had to be performed. This time the left femoral artery could not be used because of a postinterventional complication so we used the arterial cannula of the ECMO system as the access for the ablation catheter using a Y-connector. Using this way again a successful ablation procedure could be performed, after getting familiar with manipulation the ablation catheter over the ECMO cannula and with the help of different curved ablation catheters. The issue of compromising of the effective lumen of the arterial cannula by the ablation catheter`s cross sectional area could be overcome with increasing the rotational speed of the V-A ECMO.ConclusionAblation of ventricular arrhythmias using a Y-connector to insert the ablation catheter into the arterial cannula is feasible in patients with a V-A ECMO system avoiding additional arterial puncture with potentially major vascular complications in critically ill patients. Manipulation of the catheter is not as easy as using a standard sheath but can well be performed after a short habituation.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.