    RF tumor ablation with internally cooled electrodes and saline infusion: what is the optimal location of the saline infusion?

    • Fernando Burdío, Enrique J Berjano, Ana Navarro, José M Burdío, Antonio Güemes, Luis Grande, Ramón Sousa, Jorge Subiró, Ana Gonzalez, Ignacio Cruz, Tomás Castiella, Eloy Tejero, Ricardo Lozano, and Miguel A de Gregorio.
    • Department of Surgery, Hospital del Mar, Barcelona, Spain. fburdio@hotmail.com
    • Biomed Eng Online. 2007 Jan 1;6:30.

    BackgroundRadiofrequency ablation (RFA) of tumors by means of internally cooled electrodes (ICE) combined with interstitial infusion of saline may improve clinical results. To date, infusion has been conducted through outlets placed on the surface of the cooled electrode. However, the effect of infusion at a distance from the electrode surface is unknown. Our aim was to assess the effect of perfusion distance (PD) on the coagulation geometry and deposited power during RFA using ICE.MethodsExperiments were performed on excised bovine livers. Perfusion distance (PD) was defined as the shortest distance between the infusion outlet and the surface of the ICE. We considered three values of PD: 0, 2 and 4 mm. Two sets of experiments were considered: 1) 15 ablations of 10 minutes (n > or = 4 for each PD), in order to evaluate the effect of PD on volume and diameters of coagulation; and 2) 20 additional ablations of 20 minutes. The effect of PD on deposited power and relative frequency of uncontrolled impedance rises (roll-off) was evaluated using the results from the two sets of experiments (n > or = 7 for each PD). Comparisons between PD were performed by analysis of variance or Kruskal-Wallis test. Additionally, non-linear regression models were performed to elucidate the best PD in terms of coagulation volume and diameter, and the occurrence of uncontrolled impedance rises.ResultsThe best-fit least square functions were always obtained with quadratic curves where volume and diameters of coagulation were maximum for a PD of 2 mm. A thirty per cent increase in volume coagulation was observed for this PD value compared to other values (P < 0.05). Likewise, the short coagulation diameter was nearly twenty five per cent larger for a 2 mm PD than for 0 mm. Regarding deposited power, the best-fit least square function was obtained by a quadratic curve with a 2 mm PD peak. This matched well with the higher relative frequency of uncontrolled impedance rises for PD of 0 and 4 mm.ConclusionSaline perfusion at around 2 mm from the electrode surface while using an ICE in RFA improves deposition of energy and enlarges coagulation volume.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…