Anesthesia and analgesia · Nov 2005
Air detection performance of the level 1 H-1200 fluid and blood warmer.
- Ramachandra R Avula, Richard Kramer, and Charles E Smith.
- Department of Anesthesiology, MetroHealth Medical Center, 2500 MetroHealth Dr., MetroHealth Medical Center, Case Western Reserve University, Cleveland, OH 44109, USA.
- Anesth. Analg. 2005 Nov 1;101(5):1413-6.
AbstractWe evaluated the Level 1 H-1200 fluid warmer during simulated conditions of minor to massive air embolism. The fluids we tested were crystalloid and diluted red cells (estimated hematocrit 50%) during gravity and pressure driven flow. The air volumes tested ranged from 1-60 mL for crystalloid and 30-150 mL for red cells. No air was observed distal to the air detector and clamp during all test conditions. The device consistently alarmed and automatically shut off flow. Air was easily purged through the gas vent-filter assembly during simulated air embolism with crystalloid but not with red cell infusion. The use of ultrasonic air detection coupled with automatic shutoff is a significant safety improvement of the Level 1 H-1200 fluid and blood warmer.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.