• Reg Anesth Pain Med · Jul 2011

    Comparative Study

    Optical detection of the brachial plexus for peripheral nerve blocks: an in vivo swine study.

    • Marcus Brynolf, Micha Sommer, Adrien E Desjardins, Marjolein van der Voort, Stefan Roggeveen, Walter Bierhoff, Benno H W Hendriks, James P Rathmell, Alfons G H Kessels, Michael Söderman, and Björn Holmström.
    • Department of Anesthesiology and Intensive Care, Karolinska University Hospital, Huddinge, Stockholm, Sweden.
    • Reg Anesth Pain Med. 2011 Jul 1;36(4):350-7.

    Background And ObjectivesAccurate identification of nerves is critical to ensure safe and effective delivery of regional anesthesia during peripheral nerve blocks. Nerve stimulation is commonly used, but it is not perfect. Even when nerve stimulation is performed in conjunction with ultrasound guidance, determining when the needle tip is at the nerve target region can be challenging. In this in vivo pilot study, we investigated whether close proximity to the brachial plexus and penetration of the axillary artery can be identified with optical reflectance spectroscopy, using a custom needle stylet with integrated optical fibers.MethodsUltrasound-guided insertions to place the needle tip near the brachial plexus at the axillary level were performed at multiple locations in 2 swine, with the stylet positioned in the cannula of a 20-gauge stimulation needle. During each insertion, optical reflectance spectra were acquired with the needle tip in skeletal muscle, at the surface of muscle fascia, and at the nerve target region; confirmation of the final needle position was provided by nerve stimulation. In addition, an insertion to the lumen of the axillary artery was performed in a third swine. Differences in the spectra were quantified with lipid and hemoglobin parameters that provide contrast for optical absorption by the respective chromophores.ResultsThe transition of the needle tip from skeletal muscle to the nerve target region was associated with higher lipid parameter values (P < 0.001) and lower hemoglobin parameter values (P < 0.001). The transition of the needle tip from muscle fascia to the nerve target region was associated with higher lipid parameter values (P = 0.001). Intraluminal access of the axillary artery was associated with an elevated hemoglobin parameter.ConclusionsSpectroscopic information obtained with the optical needle is distinct from nerve stimulation and complementary to ultrasound imaging, and it could potentially allow for reliable identification of the injection site during peripheral nerve blocks.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.