• J. Cardiothorac. Vasc. Anesth. · Jun 1997

    Comparative Study

    Evaluation of the accuracy and response time of STAT-mode continuous cardiac output.

    • M A Lazor, E T Pierce, G D Stanley, J L Cass, E F Halpern, and R H Bode.
    • Department of Anesthesia and Critical Care, Beth Isreal Deaconess Medical Center, Harvard Medical School, Boston, MA 02215, USA.
    • J. Cardiothorac. Vasc. Anesth. 1997 Jun 1;11(4):432-6.

    ObjectivesThis study was conducted to compare continuous cardiac output (CCO) with bolus thermodilution cardiac output (BTD) at steady state, and to compare the response time of STAT CCO with that of trend CCO, mean arterial pressure, and mixed venous oxygen saturation [SvO2] during an acute hemodynamic change.DesignProspective study.SettingUniversity hospital.ParticipantsTwenty-nine patients undergoing cardiac surgery or liver transplantation.InterventionsSTAT and trend CCO were compared with BTD cardiac output during steady state intraoperatively and postoperatively in the intensive care unit. Ten patients, who required epicardial pacing after cardiac surgery, were studied to compare the response time of STAT CCO with that of trend CCO, mean arterial pressure, and BvO2 after a 10% to 20% increase in pacing rate.Measurements And Main ResultsA total of 108 cardiac output data sets were analyzed at steady state. Steady state was defined as stable heart rate and mean arterial pressure (+/- 5%) and stable central venous pressure (+/- 2 mmHg) measured immediately before and after each data set. Cardiac output ranged from 2.3 to 8.5 L/min. The correlation between STAT CCO and BTD was r = 0.94, and for trend CCO and BTD was r = 0.94. The bies and precision for STAT CCO versus BTD were 0.06 L/min (Cl 95%: -0.08 to 0.18) and 0.61 L/min. The bias and precision for trend CCO versus BTD were 0.06 L/min (Cl 95%: -0.04 to 0.16) and 0.49 L/min. Eleven data sets were analyzed to study response time of STAT CCO, which was defined as the first time the percent change of the mean of each variable was significantly increased from baseline. Significant increases in mean arterial pressure and SvO2 were detected after 30 seconds (2.5%, p = 0.01) and 90 seconds (2.0%, p = 0.04), respectively. A significant increase in STAT CCO was reached at 270 seconds (4.4%, p = 0.005). Trend CCO tended to increase but did not reach statistical significance within 6 minutes.ConclusionsSTAT and trend CCO are accurate and precise and show close agreement with BTD cardiac output at steady state. The faster algorithm of STAT CCO offers some advantage over trend CCO during an acute hemodynamic change. However, because of the averaging process for determining CCO, the response time of STAT CCO is slower than that of mean arterial pressure and SvO2.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…