• J Am Soc Echocardiogr · Jul 1997

    Normal echocardiographic characteristics of the Sorin Bicarbon bileaflet prosthetic heart valve in the mitral and aortic positions.

    • L Badano, R Mocchegiani, D Bertoli, G DeGaetano, L Carratino, L Pasetti, M Caudullo, A Budini, B Mannello, and G Passerone.
    • Servizio di Cardiologia, Ospedale di Nervi, Genoa, Italy.
    • J Am Soc Echocardiogr. 1997 Jul 1;10(6):632-43.

    AbstractDoppler echocardiographic characteristics of normally functioning Sorin Bicarbon prostheses were prospectively assessed in 226 consecutive patients (135 male and 91 female patients, mean age 61 +/- 10 years) with 233 valves in the mitral (n = 67) and aortic (n = 166) positions whose function was considered normal by clinical and echocardiographic evaluation. Patterns of "normal" transprosthetic leakage were assessed with transthoracic echocardiography in all valves and with transesophageal echocardiography in six selected mitral valve prostheses. For the mitral valve prostheses, we found that peak and mean gradient, as well as pressure half-time, were not significantly different in either the 25 or the 31 mm valves (median values from 15 to 10 mm Hg, from 4 to 4 mm Hg, and from 70 to 83 ms; p = Not significant for all). On transthoracic study, 12 patients (17%) with a Sorin Bicarbon valve in the mitral position showed minimal transprosthetic leakage. On transesophageal study, all patients showed a transprosthetic leakage whose spatial distribution had a complex pattern: in planes orthogonal to the leaflet axis, two to four jets arising from the hinge points and converging toward the center of the valve plane could be visualized; in planes parallel to the leaflet axis, there were three jets, the two lateral ones diverging and the central one perpendicular to the valve plane. For the aortic valve prostheses, there was a significant decrease in transprosthetic gradients and an increase in effective orifice areas as prosthesis size increased. Peak and mean gradients decreased from a median value of 25 and 13 mm Hg in the 19 mm valves to 9 and 5 mm Hg in the 29 mm valves, respectively. Effective prosthetic valve area calculated with the continuity equation increased from a median value of 0.97 cm2 for the 19 mm size valves to 3.45 cm2 for the 29 mm size. With analysis of variance, effective prosthetic aortic valve area differentiated various valve sizes (F = 40.9, p < 0.0001) better than peak (F = 10.3, p < 0.0001) or mean (F = 8.04, p < 0.0001) gradients alone did. Furthermore, effective prosthetic aortic valve area correlated better than peak and mean gradients with prosthetic size (r = 0.76, r = -0.45, and r = -0.39, respectively). On transthoracic study, 109 patients (66%) showed minimal transprosthetic leakage. These normal values, obtained in a large number of patients with normofunctioning mitral and aortic Sorin Bicarbon valves, may help to identify Sorin Bicarbon prosthesis dysfunction.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…