• Am. J. Cardiol. · Oct 1989

    Comparative Study

    Noninvasive measurement of cardiac output during surgery using a new continuous-wave Doppler esophageal probe.

    • A Kumar, S Minagoe, D Thangathurai, M Mikhail, D Novia, J F Viljoen, S H Rahimtoola, and P A Chandraratna.
    • Department of Medicine, Los Angeles County-University of Southern California Medical Center, University of Southern California School of Medicine 90033.
    • Am. J. Cardiol. 1989 Oct 1;64(12):793-8.

    AbstractThe ability of a new continuous-wave Doppler esophageal probe to measure cardiac output noninvasively during surgery under general anesthesia was tested and compared with simultaneously measured thermodilution cardiac output. A Doppler computer, calibrated for the aortic diameter and the transcutaneously measured cardiac output from the suprasternal notch, computed the Doppler cardiac output from the descending aortic blood flow velocity signal. A total of 246 paired Doppler cardiac output and thermodilution cardiac output measurements were made in 14 patients during surgery. The average thermodilution cardiac output was 5.90 +/- 3.27 (standard deviation) liters/min (range 1.20 to 19.18); the average Doppler cardiac output was 6.21 +/- 4.0 liters/min (range 2.30 to 28.20). The difference between the cardiac output measured by the 2 techniques was 1.38 +/- 2.2 liters/min (range 0.04 to 16.8). Two to 5 cardiac output measurements were averaged and arranged into "time periods." The average standard deviations for thermodilution and Doppler cardiac outputs within each time period were 0.64 and 0.47 liters/min, respectively. There was a correlation between the 2 measurements over a range of cardiac output values (r = 0.76, Doppler cardiac output = 0.93 x thermodilution cardiac output +0.7, standard error of the estimate = 1.76). Reproducible measurements of Doppler cardiac output were obtained during intraobserver (mean difference 0.64 +/- 0.52 liter/min) and interobserver (mean difference 0.41 +/- 0.36 liter/min) studies (n = 8). Cardiac output measurement by the Doppler esophageal probe could be used for hemodynamic monitoring during surgery in selected patients with cardiopulmonary disease.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.