The Australian Incident Monitoring Study. The oxygen analyser: applications and limitations--an analysis of 200 incident reports.
- L Barker, R K Webb, W B Runciman, and J H Van der Walt.
- Department of Anaesthesia and Intensive Care, University of Adelaide, S.A.
- Anaesth Intensive Care. 1993 Oct 1;21(5):570-4.
AbstractThe first 2000 incidents reported to the Australian INcident Monitoring Study were analysed with respect to the role of the oxygen analyser; 27 (1%) were first detected by the oxygen analyser. All of these were amongst the 1256 incidents which occurred in association with general anaesthesia, of which 48% were "human detected" and 52% "monitor detected". The oxygen analyser was ranked 7th and detected 4% of these monitor detected incidents. This figure would have been much higher had the oxygen analyser been correctly used on more occasions. The oxygen analyser detected 10 ventilator-driving-gas leaks into the circuit, 6 hypoxic mixtures due to rotameter settings, 3 inappropriate nitrous oxide concentrations, 2 disconnections and 1 leak at the common gas outlet, and 2 partial and 1 total failure of ventilation. In a theoretical analysis of these 1256 incidents it was considered that the oxygen analyser, used on its own, would have detected 114 (9%), had they been allowed to evolve (3% before any potential for organ damage). In 4 incidents an oxygen analyser gave faulty readings, in 3 caused a leak and in one a total circuit obstruction; 5 incidents were not detected because the alarm had been disabled. Despite the advent of piped gas supplies, failure of gas delivery or delivery of a "wrong" gas mixture still occurs surprisingly frequently in current anaesthetic practice; hypoxic mixtures were supplied on 16 occasions, other "wrong" mixtures on 23 and the oxygen supply failed on 7 occasions.(ABSTRACT TRUNCATED AT 250 WORDS)
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.