Accuracy and clinical performance of a continuous intra-arterial blood-gas monitoring system during thoracoscopic surgery.
- A Zollinger, D R Spahn, T Singer, M P Zalunardo, S Stoehr, W Weder, and T Pasch.
- Department of Surgery, University Hospital Zurich, Switzerland.
- Br J Anaesth. 1997 Jul 1;79(1):47-52.
AbstractAccuracy and performance of the only currently available intra-arterial blood-gas monitoring system (Paratrend 7, PT7) were assessed in 23 patients during thoracoscopic surgery using one-lung ventilation. Over a wide range of values for arterial PO2 (6.1-61.1 kPa), PCO2 (4.1-9.5 kPa) and pH (7.19-7.50), 138 arterial blood-gas values obtained by PT7 were compared with corresponding in vitro laboratory blood-gas measurements. We found good clinical performance with the PT7 and good agreement between PT7 values and in vitro measurements for arterial PO2 (bias (1.96 SD) = 0.38 (9.52) kPa), PCO2 (0.31 (0.76) kPa) and pH (-0.017 (0.065)). Also, the bias for sequential changes between two, consecutive times was not significantly different from the ideal value of 0. We conclude that the PT7 is helpful in monitoring patients during thoracoscopy.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.