-
- Giorgio Ferriero, Stefano Vercelli, Ludovit Salgovic, Valeria Stissi, and Francesco Sartorio.
- Fondazione Salvatore Maugeri, Istituto Scientifico di Veruno, Servizio di Fisiatria Occupazionale ed Ergonomia, Via per Revislate 13, I-28010 Veruno, NO, Italy. giorgio.ferriero@fsm.it
- Phys Ther. 2010 May 1;90(5):776-83.
Background And PurposeScarring after surgery can lead to a wide range of disorders. At present, the degree of scar adhesion is assessed manually and by ordinal scales. This article describes a new device (the Adheremeter) to measure scar adhesion and assesses its validity, reliability, and sensitivity to change.DesignThis was a reliability and validity study.SettingThe study was conducted at the Scientific Institute of Veruno.Participants And MethodsTwo independent raters, a physical therapist and a physical therapist student, used the Adheremeter to measure scar mobility and contralateral normal skin in a sample of 25 patients with adherent postsurgical scars before (T1) and after (T2) physical therapy. Two indexes of scar mobility, the adherence's surface mobility index (SM(A)) and the adherence severity index (AS), were calculated. Their correlation with the Vancouver Scar Scale (VSS) and its pliability subscale (PL-VSS) was assessed for the validity analysis.ResultsBoth the SM(A) and the AS showed good-to-excellent intrarater reliability (intraclass correlation coefficient [ICC]=.96) and interrater reliability (SM(A): ICC=.97 and .99; AS: ICC=.87 and .87, respectively, at T1 and T2), correlated moderately with the VSS and PL-VSS only at T1 (r(s)=-.58 to -.66), and were able to detect changes (physical therapist/physical therapist student): z score=-4.09/-3.88 for the SM(A) and -4.32/-4.24 for the AS; effect size=0.6/0.4 for the SM(A) and 1.4/1.2 for the AS; standard error of measurement=4.59/4.79 mm(2) for the SM(A) and 0.05/0.06 for the AS; and minimum detectable change=12.68/13.23 mm(2) for the SM(A) and 0.14/0.17 for the AS.LimitationsThe measurement is based on the rater's evaluation of force to stretch the skin and on the patient's judgment of comfort.Discussion And ConclusionThe Adheremeter showed a good level of reliability, validity, and sensitivity to change. Further studies are needed to confirm these results in larger cohorts and to assess the device's validity for other types of scars.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.