Randomized Controlled Trial
Evaluation of an intervention to maintain endotracheal tube cuff pressure within therapeutic range.
- Mary Lou Sole, Xiaogang Su, Steve Talbert, Daleen Aragon Penoyer, Samar Kalita, Edgar Jimenez, Jeffery E Ludy, and Melody Bennett.
- University of Central Florida in Orlando, 32826, USA. msole@mail.ucf.edu
- Am. J. Crit. Care. 2011 Mar 1;20(2):109-17; quiz 118.
BackgroundEndotracheal tube cuff pressure must be kept within an optimal range that ensures ventilation and prevents aspiration while maintaining tracheal perfusion.ObjectivesTo test the effect of an intervention (adding or removing air) on the proportion of time that cuff pressure was between 20 and 30 cm H(2)O and to evaluate changes in cuff pressure over time.MethodsA repeated-measure crossover design was used to study 32 orally intubated patients receiving mechanical ventilation for two 12-hour shifts (randomized control and intervention conditions). Continuous cuff pressure monitoring was initiated, and the pressure was adjusted to a minimum of 22 cm H(2)O. Caregivers were blinded to cuff pressure data, and usual care was provided during the control condition. During the intervention condition, cuff pressure alarm or clinical triggers guided the intervention.ResultsMost patients were men (mean age, 61.6 years). During the control condition, 51.7% of cuff pressure values were out of range compared with 11.1% during the intervention condition (P < .001). During the intervention, a mean of 8 adjustments were required, mostly to add air to the endotracheal tube cuff (mean 0.28 [SD, 0.13] mL). During the control condition, cuff pressure decreased over time (P < .001).ConclusionsThe intervention was effective in maintaining cuff pressure within an optimal range, and cuff pressure decreased over time without intervention. The effect of the intervention on outcomes such as ventilator-associated pneumonia and tracheal damage requires further study.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.