J Cardiothorac Anesth · Feb 1989
Randomized Controlled Trial Comparative Study Clinical TrialComparison among manual, computer-assisted, and closed-loop control of blood pressure after cardiac surgery.
- C J Murchie and G N Kenny.
- University Department of Anaesthesia, Glasgow Royal Infirmary, Scotland, United Kingdom.
- J Cardiothorac Anesth. 1989 Feb 1;3(1):16-9.
AbstractForty-five patients who required vasodilator therapy for systemic arterial hypertension following cardiac surgery were randomly allocated to receive: (1) manual control of a sodium nitroprusside (SNP) infusion; (2) computer-assisted control where the nurse was provided with a color graphical display of performance; or (3) automatic closed-loop control. Limits of acceptable systolic pressure were prescribed for each patient, and the percentage time spent outside these limits was calculated for each patient. The closed-loop computer system (CLCS) produced significantly better control than either manual or computer-assisted systems for the time spent outside the prescribed limits (P less than 0.001), and better control for the time spent outside the limits +/- 10 mmHg (manual P less than 0.001; assisted P less than 0.01). While the CLCS proved more satisfactory than manual control for the limits +/- 20 mmHg and +/- 30 mmHg (+/- 20 mmHg P less than 0.001; +/- 30 mmHg P less than 0.01), there was no significant difference in the quality of control between the CLCS and computer-assisted systems for these limits. The results suggest that the quality of blood pressure control achieved by nurses can be improved by providing them with a clear graphical display of their performance. Better control of blood pressure should reduce the adverse effects of hypotension and hypertension following cardiac surgery.
Notes
Knowledge, pearl, summary or comment to share?