Acta Neurochir. Suppl. · Jan 2012
Is P(br)O (2) pressure reactivity index (ORx) dependent on the type of oxygen probe? An in vivo study.
- G Grözinger, Martin Schenk, Christian Thiel, Karolin Thiel, Matthias H Morgalla, and Martin U Schuhmann.
- Department of General and Transplantation Surgery, Eberhard-Karls-University Hospital, Tübingen, Germany.
- Acta Neurochir. Suppl. 2012 Jan 1;114:173-6.
ObjectiveTo evaluate if ORx is dependent on the type of brain tissue O(2) (P(br)O(2)) probe in an in vivo setting.MethodsIn eight German landrace pigs two types of probes were implanted simultaneously in the same cerebral hemisphere. All pigs underwent hepatectomy and received neuromonitoring until death. A LICOX(®) probe CCI.S, representing a Clarke type electrode, was compared with a Raumedic Neurovent PTO, representing an optode. Data were sampled at 50 Hz. Average values were calculated every 30 s. Cerebral perfusion pressure (CPP) was averaged over 30 s. ORx was calculated for each probe. To increase the signal to noise ratio of the ORx, the ORx values, which had been assessed every minute, were averaged over 1 h.ResultsThe overall measurement time was 145.1 h (8,703 data pairs). Despite a mean difference of 6.2 mmHg (p < 0.0001) in the measured values of P(br)O(2), the mean ORx(licox) was 0.139, mean ORx(raumedic) 0.146 (p = 0.2098). Correlation coefficient of ORx values assessed every minute and every hour was 0.52 and 0.58 respectively.ConclusionDespite this significant difference in absolute values of P(br)O(2) the derived mean ORx values were not different. Similar to the established Licox system, the Raumedic system seems to enable a valid ORx recording.