Comparative Study
Portable ultrasonography and bladder volume accuracy--a comparative study using three-dimensional ultrasonography.
- Khurshid R Ghani, James Pilcher, David Rowland, Uday Patel, Daruish Nassiri, and Ken Anson.
- Department of Urology, St. George's Hospital Medical School, London, United Kingdom. krghani@gmail.com
- Urology. 2008 Jul 1;72(1):24-8.
ObjectivesTo compare the ultrasound bladder volume accuracy and level of agreement between two portable bladder scanners (Bladderscan and Bardscan) and a three-dimensional ultrasound (3D-US) system.MethodsA total of 50 healthy volunteers were scanned using the Bladderscan BVI 3000, Bardscan, and 3D-US system (HDI 4000), in random sequence. The BVI3000 is a dedicated bladder volume calculator, and the Bardscan combines real-time ultrasonography with bladder volume calculation. The ultrasound bladder volumes were compared with the voided volume measurements. The volunteers underwent repeat scanning after voiding, and those with a measurable residual volume were excluded from the final analysis.ResultsA residual volume was detected in 16 subjects (32%). In the remaining 34 subjects, the mean voided volume +/- standard deviation was 252.9 +/- 167.4 mL (range 33 to 709). A significant correlation (P <0.001) was found between the voided and ultrasound volumes with all three methods (Bardscan, r = 0.97; Bladderscan, r = 0.98; and 3D-US system, r = 0.99). No significant differences were found between the voided volumes and the Bladderscan or 3D-US volumes; however, the Bardscan significantly underestimated the voided volume by a mean of 21.4 mL (t = 2.84, P = 0.0076). The Bland-Altman 95% limit of agreement between the voided and calculated volumes was -64.5 to 107.2 mL, -73.7 to 88.4 mL, and -28.9 to 40.0 mL for the Bardscan, Bladderscan, and 3D-US systems, respectively.ConclusionsThe results of our study have shown that although the Bardscan has the advantages of real-time scanning with portability and instantaneous volume calculation, it is not as accurate as the Bladderscan. The accuracy and level of clinical agreement was greatest when using the 3D-US system to calculate the bladder volume.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.